Product description
Whether in the field or in the laboratory,this analyzers are leaders in portable rapid analysis instruments. We designs and produces analytical instruments for petrol/gasoline, diesel fuel, kerosene.
Portable octane and cetane analyzers are often used for on-site analysis of octane number of power gasoline, corresponding to the motor method and research method (RON and MON), and also suitable for cetane number analysis of diesel. In addition, the temperature of the analyzed liquid and the freezing point of diesel can also be measured, and the measurement results are displayed on the LCD liquid crystal display.
Octane and cetane analyzer is widely used all over the world. Its measurement method conforms to international standards: octane measurement conforms to: ASTM D 2699-86, ASTM D 2700-86; diesel cetane measurement conforms to: ASTM D 4737- 03, ASTM D 613, EN ISO 5165.
Function
1. Determining octane numbers for motor petrols;
2. Determining cetane numbers for diesel fuels;
3. Determining solidification temperature and diesel fuel type;
4. Content of kerosene in diesel fuel;
5. Petrol breakdown time (oxidative stability);
Measuring principle
The principle of Octane Meter operation is in determining knock characteristic of gasoline, self-flammability of diesel fuel based on measurements of their dielectric permeability.
Technical specifications
# | Parameter description | Measurement units | Value |
1 | Measured petrols octane numbers range | ON | 40-135 |
2 | Acceptable limit of octane number measurement basic error, max | ON | ±0.5 |
3 | Limit of acceptable difference between parallel octane number measurements, max | ON | ±0.2 |
4 | Petrol oxidation breakdown time measurement range | Min | 50-2400 |
5 | Acceptable basic error limit of petrol oxidation breakdown time | min | 10 |
6 | Cetane numbers measurement range | CN | 20-100 |
7 | Acceptable basic error limit of cetane numbers, max | CN | ±1.0 |
8 | Acceptable difference limit between cetane numbers parallel measurements, max |
CN |
± 0.5 |
9 | Acceptable error limit when determining diesel-fuel pour point | ºC | 5 |
10 | Kerosene content determination range in diesel fuels | % | 0-95 |
11 | Acceptable basic error limit when determining kerosene content in diesel fuels |
% |
3 |
12 | Measurement time | sec | 1–5 |
13 | Insufficient power supply indication operation threshold | V | 5.4 |
14 | Instrument useful life | Minimum years | 6 |
15 | Overall dimensions of: | ||
electronic module | mm | 100×210×40 | |
sensor #1 | mm | 60×100 | |
Instrument weight | g | 680 |
Home Products About Us News FAQ Factory Show Download Contact Us
No. 1, Jianlong N. Rd., Jiulongpo District,Jiulong Industrial Park, Chongqing, China
sales@topoilpurifier.com
+86-13983391036
topoilpurifier
+86-13983391036