Application:
This instrument is designed and manufactured under the standard SH / T0175 in “The Determination of Distillate Fuel Oxidation Stability," but also comply with ASTM D2274. Applicable to the determination of distillate fuel oxidation stability by accelerating method.
Technical parameters:
1. Heating power: control of heating: 800W auxiliary heating: 1500W
2. Temperature range: room temperature ~ 100 ℃
3. The working temperature: 95 ± 0.2 ℃
4. Glass rotameter flow: 50 ± 5mL / min (4 sticks)
5. The flow controller performance: input flow rate variation ± 10%
6. Hole number: 4 (four samples can also do) (HF-330A type 6 holes)
7. Timer: 0 ~ 99h
8. Instrument gauges: input pressure: 0.3Mpa Output pressure: 0.1Mpa
9. Power supply: AC 220v ± 10% 50Hz
10. Dimensions: 560 * 510 * 1400mm (L*W*H)
11. The membrane: cellulose ester film, diameter 47 ~ 50mm, aperture 0.8μm stable physical and chemical properties.
12. Dimensions: 560 *510 *1400 (mm)
13. Weight: about 40kg